Chemical SOP
Microbiology SOP
Warehouse SOP
Manufacturing SOP
Information technology SOP

DAILY BALANCE VERIFICATION [PICTORIAL]

Daily balance verification in pharma is the routine, documented checking of analytical balances each day they’re used to ensure they are accurate and fit for use in GMP testing. It’s a pre-use fitness check, not a full calibration.

Here’s a crisp, SOP-style explanation you can directly adapt.


1. Purpose

  • To confirm that each analytical balance used for GMP/testing is:
    • Functioning correctly
    • Within predefined accuracy/precision limits
    • Safe to use for weighing of reagents, standards, and samples

If daily verification fails → do not use the balance for any GMP work until investigated and found acceptable.


2. What Needs to Be Defined in SOP

Your “Daily Balance Verification” SOP should clearly define:

  1. Scope
    • All analytical balances used in:
      • QC lab
      • Production (dispensing, weighing rooms)
      • R&D (if data used for CGMP decisions)
  2. Frequency
    • At the beginning of each day of use
    • Additionally:
      • After relocation
      • After power failure
      • If the balance shows abnormal behavior (drift, error messages, or suspected damage)
  3. Responsibility
    • Trained analysts / production operators
    • Review by QA/QC supervisor (for trend and OOS actions)

3. Pre-Checks Before Verification

Before starting daily checks:

  • Ensure balance is level (spirit bubble centered).
  • Draft shield doors closed (if applicable).
  • Balance is clean (no spills or powder).
  • Warm-up time completed (as per manufacturer).
  • No strong air draft, vibration, or direct sunlight.

4. Standard Weights and Their Control

Use certified reference weights with known values and tolerances:

  • Typically Class E2 / F1 OIML or equivalent, depending on balance readability.
  • Maintain:
    • Weight ID (e.g., W-10g-01)
    • Nominal value (e.g., 10 g)
    • Certificate number, traceability to national/international standards
    • Calibration due date

Weights must be:

  • Stored in clean, closed boxes.
  • Handled with tweezers or gloves to avoid contamination.
  • Kept near the balance (same environment) for temperature equilibrium.

5. Typical Daily Verification Procedure

You can express this as a step-wise checklist in the SOP/poster:

  1. Power On & Warm-Up
    • Switch on balance (if not 24×7).
    • Allow recommended warm-up period.
  2. Zero / Tare Check
    • With empty pan, press ZERO.
    • Confirm reading is 0.0000 g (or 0.00000 g depending on resolution).
  3. Internal Adjustment (If Available)
    • Run internal calibration/adjustment as per manufacturer’s instructions.
    • Document performed, if your SOP requires.
  4. Place Certified Weight(s)
    Recommended approach:
    • Use at least two weights:
      • One near the lower or mid-range (e.g., 10 g)
      • One near the upper range (e.g., 100 g) of routine use
    • Record displayed value vs. nominal value.
  5. Calculate Error
    • Error = |Displayed value – Nominal value|
    • Check against predefined allowable tolerance (e.g., ±0.2 mg for 10 g, etc., based on risk and balance capability).
  6. Repeatability Check (Optional but Good Practice)
    • Weigh the same weight 3–5 times, removing and replacing it.
    • Calculate standard deviation (or range).
    • Compare with defined acceptance criteria (e.g., σ ≤ 0.15 mg).
  7. Return to Zero
    • Remove weight and ensure balance returns to 0.0000.
    • If not, investigate and repeat as needed.
  8. Documentation
    • Record in a log:
      • Date & time
      • Balance ID
      • Weight ID(s)
      • Nominal vs. observed values
      • Calculated error
      • Pass/Fail
      • Analyst initials/signature

6. Acceptance Criteria (Example)

These must be defined in your validation/QRM, but common pattern:

  • Accuracy:
    • Error for 10 g weight must be within ±0.002 g
    • Error for 100 g weight within ±0.02 g
      (Adjust for your balance readability and process criticality.)
  • Repeatability:
    • Standard deviation ≤ 0.0002 g for 10 g weight over 5 measurements
    • Or %RSD within predetermined limits

You can link these to your calibration certificate and balance qualification report.


7. What if the Balance Fails?

If any verification step fails:

  1. Immediately label the balance “OUT OF SERVICE”.
  2. Stop using it for any GMP weighing.
  3. Notify supervisor/QA.
  4. Initiate investigation:
    • Check if wrong weight used, dirty pan, drafts, etc.
    • Re-verify after corrective actions.
  5. Evaluate impact on previously weighed materials:
    • Trace back to last successful verification.
    • Assess impact on results and product quality.
    • Document in deviation/OOS and decide on re-testing or re-weighing if possible.

8. Link to Calibration and Qualification

Daily verification is complementary to:

  • Periodic external calibration (e.g., 6–12 months).
  • Equipment qualification (IQ/OQ/PQ).
  • Preventive maintenance.

Regulators expect:

  • Documented evidence that equipment is fit for intended use at the time of use.
  • Daily checks + calibration + qualification together provide that assurance.

Leave a Reply

Your email address will not be published. Required fields are marked *

error: Content is protected !!