Chemical SOP
Microbiology SOP
Warehouse SOP
Manufacturing SOP
Information technology SOP

Daily balance verification in pharma is the routine, documented checking of analytical balances each day they’re used to ensure they are accurate and fit for use in GMP testing. It’s a pre-use fitness check, not a full calibration.

Here’s a crisp, SOP-style explanation you can directly adapt.


1. Purpose

If daily verification fails → do not use the balance for any GMP work until investigated and found acceptable.


2. What Needs to Be Defined in SOP

Your “Daily Balance Verification” SOP should clearly define:

  1. Scope
    • All analytical balances used in:
      • QC lab
      • Production (dispensing, weighing rooms)
      • R&D (if data used for CGMP decisions)
  2. Frequency
    • At the beginning of each day of use
    • Additionally:
      • After relocation
      • After power failure
      • If the balance shows abnormal behavior (drift, error messages, or suspected damage)
  3. Responsibility
    • Trained analysts / production operators
    • Review by QA/QC supervisor (for trend and OOS actions)

3. Pre-Checks Before Verification

Before starting daily checks:


4. Standard Weights and Their Control

Use certified reference weights with known values and tolerances:

Weights must be:


5. Typical Daily Verification Procedure

You can express this as a step-wise checklist in the SOP/poster:

  1. Power On & Warm-Up
    • Switch on balance (if not 24×7).
    • Allow recommended warm-up period.
  2. Zero / Tare Check
    • With empty pan, press ZERO.
    • Confirm reading is 0.0000 g (or 0.00000 g depending on resolution).
  3. Internal Adjustment (If Available)
    • Run internal calibration/adjustment as per manufacturer’s instructions.
    • Document performed, if your SOP requires.
  4. Place Certified Weight(s)
    Recommended approach:
    • Use at least two weights:
      • One near the lower or mid-range (e.g., 10 g)
      • One near the upper range (e.g., 100 g) of routine use
    • Record displayed value vs. nominal value.
  5. Calculate Error
    • Error = |Displayed value – Nominal value|
    • Check against predefined allowable tolerance (e.g., ±0.2 mg for 10 g, etc., based on risk and balance capability).
  6. Repeatability Check (Optional but Good Practice)
    • Weigh the same weight 3–5 times, removing and replacing it.
    • Calculate standard deviation (or range).
    • Compare with defined acceptance criteria (e.g., σ ≤ 0.15 mg).
  7. Return to Zero
    • Remove weight and ensure balance returns to 0.0000.
    • If not, investigate and repeat as needed.
  8. Documentation
    • Record in a log:
      • Date & time
      • Balance ID
      • Weight ID(s)
      • Nominal vs. observed values
      • Calculated error
      • Pass/Fail
      • Analyst initials/signature

6. Acceptance Criteria (Example)

These must be defined in your validation/QRM, but common pattern:

You can link these to your calibration certificate and balance qualification report.


7. What if the Balance Fails?

If any verification step fails:

  1. Immediately label the balance “OUT OF SERVICE”.
  2. Stop using it for any GMP weighing.
  3. Notify supervisor/QA.
  4. Initiate investigation:
    • Check if wrong weight used, dirty pan, drafts, etc.
    • Re-verify after corrective actions.
  5. Evaluate impact on previously weighed materials:
    • Trace back to last successful verification.
    • Assess impact on results and product quality.
    • Document in deviation/OOS and decide on re-testing or re-weighing if possible.

8. Link to Calibration and Qualification

Daily verification is complementary to:

Regulators expect:

Leave a Reply

Your email address will not be published. Required fields are marked *

error: Content is protected !!