Chemical SOP
Microbiology SOP
Warehouse SOP
Manufacturing SOP
Information technology SOP

DECODING BULGING PROBLEM IN TABLET STRIPS (PICTORIAL)

1.0 Objective:

To describe the procedure for analyzing and identifying gases formed in the blister cavities of Amoxicillin and Potassium Clavulanate tablets using Gas Chromatography (GC).

2.0 Scope:

This SOP applies to the analysis of gases collected from bulged or swollen blister packs suspected to contain trapped gases due to degradation or leakage.

3.0 Responsibility:

  • QC Analyst: Perform the test and record results.
  • QC Reviewer: Verify chromatograms and calculations.
  • QA Officer: Review and approve the final report.

4.0 References:

5.0 Apparatus & Materials:

  • Gas Chromatograph with Thermal Conductivity Detector (TCD)
  • Columns: Molecular Sieve 5A and/or Porapak Q
  • Gas-tight syringe (250–500 µL, Hamilton type)
  • Helium (carrier gas, high purity)
  • Certified gas standards (CO₂, O₂, N₂)
  • Crimp vials (10–20 mL) with butyl/PTFE septa
  • Micro-needle and piercing assembly
  • Isopropyl alcohol wipes

6.0 Procedure:

6.1 Sample Collection:

  1. Select representative bulged blister packs.
  2. Clean surface with isopropyl alcohol and allow to dry.
  3. Insert a fine needle through the bulged cavity without damaging the tablet.
  4. Using a gas-tight syringe, withdraw 200–500 µL of headspace gas.
  5. Immediately inject into GC or transfer into a sealed GC vial for short-term storage (<30 min).

6.2 Instrument Setup:

ParameterSetting
DetectorTCD
Carrier GasHelium or Nitrogen, 20–30 mL/min
ColumnMolsieve 5A / Porapak Q
Injector Temp100–150 °C
Oven Temp40–60 °C (isothermal)
Detector Temp150–200 °C
Injection ModeSplitless

6.3 Calibration:

  1. Inject standard gas mixtures (low, medium, high) containing CO₂, O₂, N₂.
  2. Plot calibration curves for each gas (Area vs % concentration).
  3. Verify correlation coefficient (R² ≥ 0.995).
  4. Run a mid-level calibration verification after every 10 samples.

6.4 Sample Injection:

  1. Purge syringe once with sample gas before final draw.
  2. Inject 100–500 µL of sample directly into the GC inlet.
  3. Record chromatogram and note retention times for CO₂, O₂, N₂.

6.5 Calculations:

  • Determine gas concentration using the external calibration curve.
  • For air ingress:
    • Expected ratio: N₂:O₂ ≈ 79:21.
    • Excess CO₂ indicates possible chemical degradation (clavulanate hydrolysis).
  • Report values in % v/v or ppm.

6.6 Acceptance Criteria:

ParameterExpected RangeInterpretation
CO₂≤0.04%Normal (ambient)
CO₂ >0.5%Product degradation / moisture ingress
O₂:N₂ ratio ≈ 21:79Air ingress (seal failure)
Other gases (H₂, CH₄, etc.)Not detectedContamination if present

7.0 Quality Control:

  • Run blank and standard before and after each batch.
  • Duplicate injection RPD ≤ 10%.
  • Maintain all chromatograms and raw data with sample ID and date.

8.0 Precautions:

  • Avoid contact between the needle and the tablet.
  • Perform all gas handling in a ventilated enclosure.
  • Do not reuse syringes between samples without purging.

9.0 Documentation:

Attach chromatograms, calibration curves, calculation sheets, and investigation reports to the final record.

10.0 Interpretation and Reporting:

Based on gas composition:

  • High CO₂ → degradation due to moisture.
  • O₂/N₂ presence → seal leakage or packaging defect.
  • Other gases detected → further investigation needed.

Leave a Reply

Your email address will not be published. Required fields are marked *

error: Content is protected !!