Chemical SOP
Microbiology SOP
Warehouse SOP
Manufacturing SOP
Information technology SOP

DECODING BULGING PROBLEM IN TABLET STRIPS (PICTORIAL)

1.0 Objective:

To describe the procedure for analyzing and identifying gases formed in the blister cavities of Amoxicillin and Potassium Clavulanate tablets using Gas Chromatography (GC).

2.0 Scope:

This SOP applies to the analysis of gases collected from bulged or swollen blister packs suspected to contain trapped gases due to degradation or leakage.

3.0 Responsibility:

  • QC Analyst: Perform the test and record results.
  • QC Reviewer: Verify chromatograms and calculations.
  • QA Officer: Review and approve the final report.

4.0 References:

5.0 Apparatus & Materials:

  • Gas Chromatograph with Thermal Conductivity Detector (TCD)
  • Columns: Molecular Sieve 5A and/or Porapak Q
  • Gas-tight syringe (250–500 µL, Hamilton type)
  • Helium (carrier gas, high purity)
  • Certified gas standards (COâ‚‚, Oâ‚‚, Nâ‚‚)
  • Crimp vials (10–20 mL) with butyl/PTFE septa
  • Micro-needle and piercing assembly
  • Isopropyl alcohol wipes

6.0 Procedure:

6.1 Sample Collection:

  1. Select representative bulged blister packs.
  2. Clean surface with isopropyl alcohol and allow to dry.
  3. Insert a fine needle through the bulged cavity without damaging the tablet.
  4. Using a gas-tight syringe, withdraw 200–500 µL of headspace gas.
  5. Immediately inject into GC or transfer into a sealed GC vial for short-term storage (<30 min).

6.2 Instrument Setup:

ParameterSetting
DetectorTCD
Carrier GasHelium or Nitrogen, 20–30 mL/min
ColumnMolsieve 5A / Porapak Q
Injector Temp100–150 °C
Oven Temp40–60 °C (isothermal)
Detector Temp150–200 °C
Injection ModeSplitless

6.3 Calibration:

  1. Inject standard gas mixtures (low, medium, high) containing COâ‚‚, Oâ‚‚, Nâ‚‚.
  2. Plot calibration curves for each gas (Area vs % concentration).
  3. Verify correlation coefficient (R² ≥ 0.995).
  4. Run a mid-level calibration verification after every 10 samples.

6.4 Sample Injection:

  1. Purge syringe once with sample gas before final draw.
  2. Inject 100–500 µL of sample directly into the GC inlet.
  3. Record chromatogram and note retention times for COâ‚‚, Oâ‚‚, Nâ‚‚.

6.5 Calculations:

  • Determine gas concentration using the external calibration curve.
  • For air ingress:
    • Expected ratio: Nâ‚‚:Oâ‚‚ ≈ 79:21.
    • Excess COâ‚‚ indicates possible chemical degradation (clavulanate hydrolysis).
  • Report values in % v/v or ppm.

6.6 Acceptance Criteria:

ParameterExpected RangeInterpretation
CO₂≤0.04%Normal (ambient)
COâ‚‚ >0.5%Product degradation / moisture ingress
O₂:N₂ ratio ≈ 21:79Air ingress (seal failure)
Other gases (Hâ‚‚, CHâ‚„, etc.)Not detectedContamination if present

7.0 Quality Control:

  • Run blank and standard before and after each batch.
  • Duplicate injection RPD ≤ 10%.
  • Maintain all chromatograms and raw data with sample ID and date.

8.0 Precautions:

  • Avoid contact between the needle and the tablet.
  • Perform all gas handling in a ventilated enclosure.
  • Do not reuse syringes between samples without purging.

9.0 Documentation:

Attach chromatograms, calibration curves, calculation sheets, and investigation reports to the final record.

10.0 Interpretation and Reporting:

Based on gas composition:

  • High COâ‚‚ → degradation due to moisture.
  • Oâ‚‚/Nâ‚‚ presence → seal leakage or packaging defect.
  • Other gases detected → further investigation needed.

Leave a Reply

Your email address will not be published. Required fields are marked *

error: Content is protected !!

This is the Premium Content

You can access this page after paying the subscription fees of 21 ₹ /month only.